Olink

Olink®
Part of Thermo Fisher Scientific

Documents

Access a comprehensive collection of resources, including brochures, infographics, user manuals, validation data, technical notes, and white papers. Easily filter by document type, application, research area, and products to find the specific information you need.

Filters

DocumentsDocument typeApplicationResearch areaProduct

SDS Olink® Target 48 Control kit

SDS Olink® Target 48 Detection kit

SDS Olink® Target 48 Probe kit

SDS Olink® Target 96 Detection and Controls

SDS Olink® Target 96 Probe Kit

SDS Olink® Target96/Explore Sample Diluent

Statistical Analysis Service Agreement

For use when purchasing data analysis services.

Validation data - cardiometabolic

Panel uses a 1:2025 dilution of sample. The measuring ranges shown in Figure 3 correspond to the concentrations in the original biological sample, whereas the figures quoted in Table 1 represent the actual concentration of recombinant analyte measurable in the validation assay. A multiplication factor of 2025 should therefore be taken into consideration when comparing the addressable biological concentration to the in vitro validation data.

Validation data - cardiovascular II

Validation data - cardiovascular III

The CVD III panel uses a 1:100 dilution of sample. The measuring ranges shown in Figure 3 correspond to the concentrations in the original biological sample, whereas the figures quoted in Table 1 represent the actual concentration of recombinant analyte measurable in the validation assay. A multiplication factor of 100 should therefore be taken into consideration when comparing the addressable biological concentration to the in vitro validation data.

Validation data - cell regulation

Validation data - development

Panel uses a 1:100 dilution of sample. The measuring ranges shown in Figure 3 correspond to the concentrations in the original biological sample, whereas the figures quoted in Table 1 represent the actual concentration of recombinant analyte measurable in the validation assay. A multiplication factor of 100 should therefore be taken into consideration when comparing the addressable biological concentration to the in vitro validation data.

Validation data - immune response

Validation data - immuno-oncology

Note that this version (2.1) applies to the latest panel version with article number 95311

Validation data - inflammmation

Note that this version (3.0) applies to the latest panel version with article number 95302

Validation data - metabolism

The Metabolism panel uses a 1:10 dilution of sample. The measuring ranges shown in Figure 3 correspond to the concentrations in the original biological sample, whereas the figures quoted in Table 1 represent the actual concentration of recombinant analyte measurable in the validation assay. A multiplication factor of 10 should therefore be taken into consideration when comparing the addressable biological concentration to the in vitro validation data.

Validation data - mouse exploratory

Validation data - neuro exploratory

Note that this version of the document (1.2) applies to the latest version of the panel with article number 95391 (released July 1, 2019)

Validation data - neurology

Validation data - Olink® Flex

189 documents