Target 96 Organ Damage

Gene
ALDH3A1

Uniprot
P30838

Protein
Aldehyde dehydrogenase, dimeric NADP-preferring

See alternative names Aldehyde dehydrogenase family 3 member A1,
ALDHIII,
Aldehyde dehydrogenase 3

Uniprot Function Description

ALDHs play a major role in the detoxification of alcohol-derived acetaldehyde (Probable). They are involved in the metabolism of corticosteroids, biogenic amines, neurotransmitters, and lipid peroxidation (Probable). Oxidizes medium and long chain aldehydes into non-toxic fatty acids (PubMed:1737758). Preferentially oxidizes aromatic aldehyde substrates (PubMed:1737758). Comprises about 50 percent of corneal epithelial soluble proteins (By similarity). May play a role in preventing corneal damage caused by ultraviolet light (By similarity).

Sample type

Recommended sample types are EDTA plasma and serum. A range of additional sample types are compatible with the technology (PEA), including citrate plasma, heparin plasma, cerebrospinal fluid, (CSF), tissue and cell lysates, fine needle biopsies, microdialysis fluid, cell culture media, dried blood spots, synovial fluid, saliva, plaque extract and microvesicles. Please note that protein expression levels are expected to vary in different sample types. Certain assays are differentially affected by interfering substances such as hemolysate. Download any of our Data Validation documents or contact support@olink.com for more information.

Precision

Precision (repeatability) is calculated from linearized NPX values over LOD.

Within run precision Coefficient of Variation (CV)
5%
Between run precision Coefficient of Variation (CV)
19%

Analytical measuring range

The technical data reported below refers to the measured value in the in vitro validation assays run using known concentrations of recombinant antigen. Please note that these figures are for reference only and CANNOT be used to convert NPX values to absolute concentrations for proteins measured in plasma or serum samples.

LOD (pg/mL)
0.950
LLOQ (pg/mL)
1.9
ULOQ (pg/mL)
7812
Hook (pg/mL)
31250
Range (logs)
3.6

Sensitivity plot

The calibrator curve shown below visualizes the analytical measuring range data based on in vitro measurement of recombinant antigen. Please note that this is shown for reference only and CANNOT be used to convert NPX values to absolute concentrations for proteins measured in plasma or serum samples. The vertical dotted lines represent LLOQ and ULOQ respectively, and the horizontal line indicates the LOD.

SensitivityPlot

Biomarker Validation Data

Additional validation data, as well as a more detailed description of how the Olink panels are quality controlled can be found in our Data Validation documents – go to Document download center