Olink

Olink®
Part of Thermo Fisher Scientific

ONCOLOGY

Gene
FOLR1

Uniprot
P15328

Protein
Folate receptor alpha

See alternative names Adult folate-binding protein,
Folate receptor 1,
Folate receptor, adult,
KB cells FBP,
Ovarian tumor-associated antigen MOv18

Uniprot Function Description

Binds to folate and reduced folic acid derivatives and mediates delivery of 5-methyltetrahydrofolate and folate analogs into the interior of cells. Has high affinity for folate and folic acid analogs at neutral pH. Exposure to slightly acidic pH after receptor endocytosis triggers a conformation change that strongly reduces its affinity for folates and mediates their release. Required for normal embryonic development and normal cell proliferation.

Sample type

Human EDTA plasma and serum are the recommended sample types. Human citrate plasma and heparin plasma are also accepted. For other samples types e.g cerebrospinal fluid, (CSF), tissue or cell lysates please we recommend Olink Target 96 panels. Please note that protein expression levels are expected to vary in different sample types and certain assays may be affected by interfering substances such as hemolysate.

Precision

Precision (repeatability) is calculated from linearized NPX values over LOD.

Within run precision Coefficient of Variation (CV)
8%
Between run precision Coefficient of Variation (CV)
9%

Analytical measuring range

The technical data reported below refers to the measured value in the in vitro validation assays run using known concentrations of recombinant antigen. Please note that these figures are for reference only and CANNOT be used to convert NPX values to absolute concentrations for proteins measured in plasma or serum samples.

LOD (pg/mL)
0.8
LLOQ (pg/mL)
1.5
ULOQ (pg/mL)
3125
Hook (pg/mL)
12500
Range (logs)
3.3

Dilution factor

For optimal assay readout, Olink Explore is run using different dilutions of the original samples (undiluted, 1:10, 1:100 or 1:1000). The dilution factor for this assay is noted below and should be taken into account when estimating the appropriate addressable biological concentration of the protein based on the in vitro validation data.

Dilution factor
1:10

Sensitivity plot

The calibrator curve shown below visualizes the analytical measuring range data based on in vitro measurement of recombinant antigen. Please note that this is shown for reference only and CANNOT be used to convert NPX values to absolute concentrations for proteins measured in plasma or serum samples. The vertical dotted lines represent LLOQ and ULOQ respectively, and the horizontal line indicates the LOD.

10⁻⁵10⁻⁴10⁻³10⁻²10⁻¹10⁰10¹10²10³10⁴10⁵10⁶121086420−2−4−6−8pg/mLNPXReplicate4PLLODLLOQULOQ

Sample distribution plot

The plot below shows the levels of protein measured in a number of commercial plasma samples. Healthy subjects are shown in blue and samples obtained from patients with a range of diseases are shown in red. The latter include inflammatory, cardiovascular, autoimmune & neurological diseases, as well as cancer. The data is shown to give a general idea of the sort of data range to expect, but cannot cover all potential levels that may be seen in clinical samples.

121086420−2−4−6−8NPXControl SamplesDisease Samples