Olink

Olink®
Part of Thermo Fisher Scientific

FLEX

Gene
CCL2

Uniprot
P13500

Protein
C-C motif chemokine 2

See alternative names
Monocyte secretory protein JE,
HC11,
Monocyte chemotactic protein 1,
Monocyte chemoattractant protein 1,
Monocyte chemotactic and activating factor,
Small-inducible cytokine A2

Uniprot Function Description

Acts as a ligand for C-C chemokine receptor CCR2 (PubMed:9837883, PubMed:10587439, PubMed:10529171). Signals through binding and activation of CCR2 and induces a strong chemotactic response and mobilization of intracellular calcium ions (PubMed:9837883, PubMed:10587439). Exhibits a chemotactic activity for monocytes and basophils but not neutrophils or eosinophils (PubMed:8627182, PubMed:9792674, PubMed:8195247). May be involved in the recruitment of monocytes into the arterial wall during the disease process of atherosclerosis (PubMed:8107690).

Precision

Precision (repeatability) is calculated from linearized NPX values over LOD.

Within run precision Coefficient of Variation (CV)
7%
Between run precision Coefficient of Variation (CV)
8%

Analytical measuring range

The technical data reported below refers to the measured value in the in vitro validation assays run using known concentrations of recombinant antigen. Please note that these figures are for reference only and CANNOT be used to convert NPX values to absolute concentrations for proteins measured in plasma or serum samples.

LOD (pg/mL)
0.054
LLOQ (pg/mL)
0.135
ULOQ (pg/mL)
3224
Hook (pg/mL)
20147
Range (logs)
4.4

Sensitivity plot

The calibrator curve shown below visualizes the analytical measuring range data based on in vitro measurement of recombinant antigen. Please note that this is shown for reference only and CANNOT be used to convert NPX values to absolute concentrations for proteins measured in plasma or serum samples. The vertical dotted lines represent LLOQ and ULOQ respectively, and the horizontal line indicates the LOD.

10⁻⁴10⁻³10⁻²10⁻¹10⁰10¹10²10³10⁴10⁵181614121086420−2pg/mLNPXReplicate4PLLODLLOQULOQ

Sample distribution plot

The plot below shows the levels of protein measured in a number of commercial plasma samples. Healthy subjects are shown in blue and samples obtained from patients with a range of diseases are shown in red. The latter include inflammatory, cardiovascular, autoimmune & neurological diseases, as well as cancer. The data is shown to give a general idea of the sort of data range to expect, but cannot cover all potential levels that may be seen in clinical samples.

Biomarker Validation Data

Additional validation data, as well as a more detailed description of how the Olink panels are quality controlled can be found in our Data Validation documents – go to Document download center

Search all Biomarkers