Olink

Olink®
Part of Thermo Fisher Scientific

FLEX

Gene
TNFSF13

Uniprot
O75888

Protein
Tumor necrosis factor ligand superfamily member 13

See alternative names
A proliferation-inducing ligand,
TNF-related death ligand 1,
TNF- and APOL-related leukocyte expressed ligand 2

Uniprot Function Description

Cytokine that binds to TNFRSF13B/TACI and to TNFRSF17/BCMA. Plays a role in the regulation of tumor cell growth. May be involved in monocyte/macrophage-mediated immunological processes.

Precision

Precision (repeatability) is calculated from linearized NPX values over LOD.

Within run precision Coefficient of Variation (CV)
5%
Between run precision Coefficient of Variation (CV)
7%

Analytical measuring range

The technical data reported below refers to the measured value in the in vitro validation assays run using known concentrations of recombinant antigen. Please note that these figures are for reference only and CANNOT be used to convert NPX values to absolute concentrations for proteins measured in plasma or serum samples.

LOD (pg/mL)
18.8
LLOQ (pg/mL)
46.9
ULOQ (pg/mL)
71572
Hook (pg/mL)
1118312
Range (logs)
3.2

Sensitivity plot

The calibrator curve shown below visualizes the analytical measuring range data based on in vitro measurement of recombinant antigen. Please note that this is shown for reference only and CANNOT be used to convert NPX values to absolute concentrations for proteins measured in plasma or serum samples. The vertical dotted lines represent LLOQ and ULOQ respectively, and the horizontal line indicates the LOD.

10⁻⁵10⁻⁴10⁻³10⁻²10⁻¹10⁰10¹10²10³10⁴10⁵10⁶10⁷10⁸181614121086420−2pg/mLNPXReplicate4PLLODLLOQULOQ

Sample distribution plot

The plot below shows the levels of protein measured in a number of commercial plasma samples. Healthy subjects are shown in blue and samples obtained from patients with a range of diseases are shown in red. The latter include inflammatory, cardiovascular, autoimmune & neurological diseases, as well as cancer. The data is shown to give a general idea of the sort of data range to expect, but cannot cover all potential levels that may be seen in clinical samples.

Biomarker Validation Data

Additional validation data, as well as a more detailed description of how the Olink panels are quality controlled can be found in our Data Validation documents – go to Document download center

Search all Biomarkers