FURIN

Uniprot
P09958

Protein
Furin

See alternative names Paired basic amino acid residue-cleaving enzyme,
Dibasic-processing enzyme

Uniprot Function Description

Ubiquitous endoprotease within constitutive secretory pathways capable of cleavage at the RX(K/R)R consensus motif (PubMed:11799113, PubMed:1629222, PubMed:1713771, PubMed:2251280, PubMed:24666235, PubMed:25974265, PubMed:7592877, PubMed:7690548, PubMed:9130696). Mediates processing of TGFB1, an essential step in TGF-beta-1 activation (PubMed:7737999).

(Microbial infection) Probably cleaves and activates anthrax and diphtheria toxins.

(Microbial infection) Required for H7N1 and H5N1 influenza virus infection probably by cleaving hemagglutinin.

(Microbial infection) Able to cleave S.pneumoniae serine-rich repeat protein PsrP.

Sample type

Recommended sample types are human plasma and serum. For other sample types e.g cerebrospinal fluid, (CSF), tissue or cell lysate, please contact support@olink.com for more information. Please note that protein expression levels are expected to vary in different sample types and certain assays may be affected by interfering substances such as hemolysate.

Precision

Precision (repeatability) is calculated from linearized NPX values over LOD.

Within run precision Coefficient of Variation (CV)
11%

Analytical measuring range

The technical data reported below refers to the measured value in the in vitro validation assays run using known concentrations of recombinant antigen.

LOD (NPX)
-2.3
Hooked
Yes
Hook (NPX)
6.1

Dilution factor

For optimal assay readout, Olink assays are run using different dilutions of the original samples e.g. undiluted, 1:10, or higher. The dilution factor for this assay is noted below and should be taken into account when estimating the appropriate addressable biological concentration of the protein based on the in vitro validation data.

Dilution factor
1:10

Biomarker Validation Data

Additional validation data, as well as a more detailed description of how the Olink panels are quality controlled can be found in our Data Validation documents – go to Document download center