Olink

Olink®
Part of Thermo Fisher Scientific

Explore HT

Gene
TPX2

Uniprot
Q9ULW0

Protein
Targeting protein for Xklp2

See alternative names Differentially expressed in cancerous and non-cancerous lung cells 2,
Hepatocellular carcinoma-associated antigen 519,
Hepatocellular carcinoma-associated antigen 90,
Protein fls353,
Restricted expression proliferation-associated protein 100

Uniprot Function Description

Spindle assembly factor required for normal assembly of mitotic spindles. Required for normal assembly of microtubules during apoptosis. Required for chromatin and/or kinetochore dependent microtubule nucleation. Mediates AURKA localization to spindle microtubules (PubMed:18663142, PubMed:19208764). Activates AURKA by promoting its autophosphorylation at 'Thr-288' and protects this residue against dephosphorylation (PubMed:18663142, PubMed:19208764). TPX2 is inactivated upon binding to importin-alpha (PubMed:26165940). At the onset of mitosis, GOLGA2 interacts with importin-alpha, liberating TPX2 from importin-alpha, allowing TPX2 to activates AURKA kinase and stimulates local microtubule nucleation (PubMed:26165940).

Sample type

Recommended sample types are human plasma and serum. For other sample types e.g cerebrospinal fluid, (CSF), tissue or cell lysate, please contact support@olink.com for more information. Please note that protein expression levels are expected to vary in different sample types and certain assays may be affected by interfering substances such as hemolysate.

Precision

Precision (repeatability) is calculated from linearized NPX values over LOD.

Within run precision Coefficient of Variation (CV)
N/A

Analytical measuring range

The technical data reported below refers to the measured value in the in vitro validation assays run using known concentrations of recombinant antigen.

LOD (NPX)
2.3
Hooked
Yes
Hook (NPX)
12.6

Dilution factor

For optimal assay readout, Olink assays are run using different dilutions of the original samples e.g. undiluted, 1:10, or higher. The dilution factor for this assay is noted below and should be taken into account when estimating the appropriate addressable biological concentration of the protein based on the in vitro validation data.

Dilution factor
1:1

Biomarker Validation Data

Additional validation data, as well as a more detailed description of how the Olink panels are quality controlled can be found in our Data Validation documents – go to Document download center

Search all Biomarkers