Olink

Olink®
Part of Thermo Fisher Scientific

Target 96 Cardiovascular III

Gene
PDGFA

Uniprot
P04085

Protein
Platelet-derived growth factor subunit A

See alternative names PDGF-1,
Platelet-derived growth factor A chain,
Platelet-derived growth factor alpha polypeptide

Uniprot Function Description

Growth factor that plays an essential role in the regulation of embryonic development, cell proliferation, cell migration, survival and chemotaxis. Potent mitogen for cells of mesenchymal origin. Required for normal lung alveolar septum formation during embryogenesis, normal development of the gastrointestinal tract, normal development of Leydig cells and spermatogenesis. Required for normal oligodendrocyte development and normal myelination in the spinal cord and cerebellum. Plays an important role in wound healing. Signaling is modulated by the formation of heterodimers with PDGFB (By similarity).

Sample type

Recommended sample types are EDTA plasma and serum. A range of additional sample types are compatible with the technology (PEA), including citrate plasma, heparin plasma, cerebrospinal fluid, (CSF), tissue and cell lysates, fine needle biopsies, microdialysis fluid, cell culture media, dried blood spots, synovial fluid, saliva, plaque extract and microvesicles. Please note that protein expression levels are expected to vary in different sample types. Certain assays are differentially affected by interfering substances such as hemolysate. Download any of our Data Validation documents or contact support@olink.com for more information.

Precision

Precision (repeatability) is calculated from linearized NPX values over LOD.

Within run precision Coefficient of Variation (CV)
9%
Between run precision Coefficient of Variation (CV)
15%

Analytical measuring range

The technical data reported below refers to the measured value in the in vitro validation assays run using known concentrations of recombinant antigen. Please note that these figures are for reference only and CANNOT be used to convert NPX values to absolute concentrations for proteins measured in plasma or serum samples.

LOD (pg/mL)
0.950
LLOQ (pg/mL)
1.9
ULOQ (pg/mL)
15625
Hook (pg/mL)
31250
Range (logs)
3.9

Sensitivity plot

The calibrator curve shown below visualizes the analytical measuring range data based on in vitro measurement of recombinant antigen. Please note that this is shown for reference only and CANNOT be used to convert NPX values to absolute concentrations for proteins measured in plasma or serum samples. The vertical dotted lines represent LLOQ and ULOQ respectively, and the horizontal line indicates the LOD.

SensitivityPlot

Biomarker Validation Data

Additional validation data, as well as a more detailed description of how the Olink panels are quality controlled can be found in our Data Validation documents – go to Document download center

Search all Biomarkers