Target 48 Immune Surveillance

Gene
CCL20

Uniprot
P78556

Protein
C-C motif chemokine 20

See alternative names Small-inducible cytokine A20,
Macrophage inflammatory protein 3 alpha,
Liver and activation-regulated chemokine,
CC chemokine LARC,
Beta-chemokine exodus-1

Uniprot Function Description

Acts as a ligand for C-C chemokine receptor CCR6. Signals through binding and activation of CCR6 and induces a strong chemotactic response and mobilization of intracellular calcium ions (PubMed:11352563, PubMed:11035086, PubMed:20068036). The ligand-receptor pair CCL20-CCR6 is responsible for the chemotaxis of dendritic cells (DC), effector/memory T-cells and B-cells and plays an important role at skin and mucosal surfaces under homeostatic and inflammatory conditions, as well as in pathology, including cancer and various autoimmune diseases (PubMed:21376174). CCL20 acts as a chemotactic factor that attracts lymphocytes and, slightly, neutrophils, but not monocytes (PubMed:9038201, PubMed:11352563). Involved in the recruitment of both the proinflammatory IL17 producing helper T-cells (Th17) and the regulatory T-cells (Treg) to sites of inflammation. Required for optimal migration of thymic natural regulatory T cells (nTregs) and DN1 early thymocyte progenitor cells (By similarity). C-terminal processed forms have been shown to be equally chemotactically active for leukocytes (PubMed:11035086). Positively regulates sperm motility and chemotaxis via its binding to CCR6 which triggers Ca2+ mobilization in the sperm which is important for its motility (PubMed:23765988, PubMed:25122636). Inhibits proliferation of myeloid progenitors in colony formation assays (PubMed:9129037). May be involved in formation and function of the mucosal lymphoid tissues by attracting lymphocytes and dendritic cells towards epithelial cells (By similarity). Possesses antibacterial activity towards E.coli ATCC 25922 and S.aureus ATCC 29213 (PubMed:12149255).

Sample type

Recommended sample types are plasma and serum. A range of additional sample types are compatible with the technology (PEA), including citrate plasma, heparin plasma, cerebrospinal fluid, (CSF), tissue and cell lysates, fine needle biopsies, microdialysis fluid, cell culture media, dried blood spots, synovial fluid, saliva, plaque extract and microvesicles. Please note that protein expression levels are expected to vary between different sample types. Certain assays are differentially affected by interfering substances such as hemolysate. Download any of our Data Validation documents or contact support@olink.com for more information.

Precision

Precision (repeatability) is calculated from quantified data (pg/mL) within the limits of quantification: LOQ.

Within run precision Coefficient of Variation (CV)
5%
Between run precision Coefficient of Variation (CV)
6%

Analytical measuring range

The technical data reported below refers to the measured values generated during the assay validation process using recombinant antigens. Per default, Olink Target 48 products deliver data in absolute protein concentrations (pg/mL) using the Calibrator built into the system.

LOD (pg/mL)
0.040
LLOQ (pg/mL)
0.130
ULOQ (pg/mL)
1603
Hook (pg/mL)
8551
Range (logs)
4

Sensitivity plot

The calibrator curve shown below visualizes the analytical measuring range determined during the validation process using recombinant antigen. Please note that this is shown for reference only and CANNOT be used to e.g. convert NPX values to absolute concentrations. The vertical dotted lines represent lower limit of quantification (LLOQ) and upper limit of quantification (ULOQ) respectively. The solid horizontal line indicates the limit of detection (LOD).

SensitivityPlot

Sample distribution plot

The plot below shows the levels of protein measured in a number of plasma samples (for further information please see the Validation Data Document which can be found in the Document download center). The data is shown to give an idea of the data range to expect but cannot cover protein levels that may be seen under all potential conditions.

SampleDistributionPlot

Biomarker Validation Data

Additional validation data, as well as a more detailed description of how the Olink panels are quality controlled can be found in our Data Validation documents – go to Document download center