Target 48 Mouse Cytokine

Gene
Il31

Uniprot
Q6EAL8

Protein
Interleukin-31

Uniprot Function Description

Activates STAT3 and possibly STAT1 and STAT5 through the IL31 heterodimeric receptor composed of IL31RA and OSMR (PubMed:15184896). May function in skin immunity (PubMed:15184896). Enhances myeloid progenitor cell survival in vitro (PubMed:17379091). Induces RETNLA and serum amyloid A protein expression in macrophages (PubMed:25847241).

Sample type

Recommended sample types are plasma and serum. A range of additional sample types are compatible with the technology (PEA), including citrate plasma, heparin plasma, cerebrospinal fluid, (CSF), tissue and cell lysates, fine needle biopsies, microdialysis fluid, cell culture media, dried blood spots, synovial fluid, saliva, plaque extract and microvesicles. Please note that protein expression levels are expected to vary between different sample types. Certain assays are differentially affected by interfering substances such as hemolysate. Download any of our Data Validation documents or contact support@olink.com for more information.

Precision

Precision (repeatability) is calculated from quantified data (pg/mL) within the limits of quantification: LOQ.

Within run precision Coefficient of Variation (CV)
5%
Between run precision Coefficient of Variation (CV)
5%

Analytical measuring range

The technical data reported below refers to the measured values generated during the assay validation process using recombinant antigens. Per default, Olink Target 48 products deliver data in absolute protein concentrations (pg/mL) using the Calibrator built into the system.

LOD (pg/mL)
2.2
LLOQ (pg/mL)
2.2
ULOQ (pg/mL)
76125
Hook (pg/mL)
160000
Range (logs)
4.4

Sensitivity plot

The calibrator curve shown below visualizes the analytical measuring range determined during the validation process using recombinant antigen. Please note that this is shown for reference only and CANNOT be used to e.g. convert NPX values to absolute concentrations. The vertical dotted lines represent lower limit of quantification (LLOQ) and upper limit of quantification (ULOQ) respectively. The solid horizontal line indicates the limit of detection (LOD).

SensitivityPlot

Sample distribution plot

The plot below shows the levels of protein measured in a number of plasma samples (for further information please see the Validation Data Document which can be found in the Document download center). The data is shown to give an idea of the data range to expect but cannot cover protein levels that may be seen under all potential conditions.

SampleDistributionPlot

Biomarker Validation Data