Olink

Olink®
Part of Thermo Fisher Scientific

Target 96 Mouse Exploratory

Gene
Epcam

Uniprot
Q99JW5

Protein
Epithelial cell adhesion molecule

See alternative names Protein 289A,
Epithelial glycoprotein 314,
Tumor-associated calcium signal transducer 1

Uniprot Function Description

May act as a physical homophilic interaction molecule between intestinal epithelial cells (IECs) and intraepithelial lymphocytes (IELs) at the mucosal epithelium for providing immunological barrier as a first line of defense against mucosal infection. Plays a role in embryonic stem cells proliferation and differentiation. Up-regulates the expression of FABP5, MYC and cyclins A and E (By similarity).

Sample type

Recommended sample types are plasma and serum. A range of additional sample types are compatible with the technology (PEA), including citrate plasma, heparin plasma, cerebrospinal fluid, (CSF), tissue and cell lysates, fine needle biopsies, microdialysis fluid, cell culture media, dried blood spots, synovial fluid, saliva, plaque extract and microvesicles. Please note that protein expression levels are expected to vary between different sample types. Certain assays are differentially affected by interfering substances such as hemolysate. Download any of our Data Validation documents or contact support@olink.com for more information.

Precision

Precision (repeatability) is calculated from quantified data (pg/mL) within the limits of quantification: LOQ.

Within run precision Coefficient of Variation (CV)
8%
Between run precision Coefficient of Variation (CV)
12%

Analytical measuring range

The technical data reported below refers to the measured values generated during the assay validation process using recombinant antigens. Per default, Olink Target 48 products deliver data in absolute protein concentrations (pg/mL) using the Calibrator built into the system.

LOD (pg/mL)
7.6
LLOQ (pg/mL)
15.0
ULOQ (pg/mL)
7812
Hook (pg/mL)
7812
Range (logs)
2.7

Sensitivity plot

The calibrator curve shown below visualizes the analytical measuring range determined during the validation process using recombinant antigen. Please note that this is shown for reference only and CANNOT be used to e.g. convert NPX values to absolute concentrations. The vertical dotted lines represent lower limit of quantification (LLOQ) and upper limit of quantification (ULOQ) respectively. The solid horizontal line indicates the limit of detection (LOD).

SensitivityPlot

Biomarker Validation Data

Document download center

Search all Biomarkers