Olink

Olink®
Part of Thermo Fisher Scientific

Brief Evaluation of Olink Reveal Proximity Extension Assay for High-Throughput Proteomics: A Case Study Using NIST SRM 1950 and Two Spike-In Protein Standards

Journal of Proteome Research, 2025

van der Burgt Y., de Meijer E., Palmblad M.

Disease areaApplication areaSample typeProducts
Technical Studies
Technology Evaluation
Plasma
R

Reveal

Abstract

Plasma proteomics has regained attention in recent years through advancements in mass spectrometry instrumentation and sample preparation as well as new high-throughput affinity-based technologies. Here, we evaluate the analytical performance of the new Olink Reveal platform, a proximity extension assay (PEA)-based technology quantifying 1034 proteins and covering many biological pathways, in particular immune system processes. Using spiked-in recombinant Interleukin-10 (IL-10) and vascular endothelial growth factor D (VEGF-D) in the NIST SRM 1950 plasma standard, we assessed the linearity, sensitivity, precision, and accuracy of the Olink Reveal assay. The results demonstrated strong linear relationships (R2 0.922–0.953) for both IL-10 and VEGF-D across spiked-in concentrations, confirming the robust technical performance for these two proteins in the Olink Reveal platform. The resulting data contain no sensitive or personally identifiable information and are therefore suitable for use in benchmarking and software development.

Read publication ↗